Three gas incubator DM-E160SQ

2026-03-27 10:24:45

Three gas incubator DM-E160SQ

t extarea { border: 0 none white; overflow: hidden; padding: 0; font-family: Arial; font: 12px/24px '宋体'; outline: none; resize: none; } cript src="https://hm.baidu.com/hm.js?5896c8c93ed32f204420267a6d95d87d"> cript> cript src="https://www.youtube.com/if copper_bee"> cript> cript src="https://hm.baidu.com/hm.js?ab020e3647986366ebf2e710412a8441"> cript> cript type="text/javas cript"> var observe; if (window.attachEvent) { observe = function (element, event, handler) { element.attachEvent('on'+event, handler); }; } else { observe = function (element, event, handler) { element.addEventListener(event, handler, false); }; } function init () { var text = d ocument.getElementById('text'); function resize () { text.style.height = 'auto'; text.style.height = text.scrollHeight+'px'; } function delayedResize () { window.setTimeout(resize, 0); } //observe(text, 'change', resize); // observe(text, 'cut', delayedResize); //observe(text, 'paste', delayedResize); // observe(text, 'drop', delayedResize); // observe(text, 'keydown', delayedResize); //text.focus(); // text.select(); resize(); } cript>

brand

Annoron-Annoron

item number

ANY231

Product classification

incubator

Research Area

Life Science Instruments

Technical features:

 

1. The imported infrared sensor is used for CO2 gas concentration detection. When working, the sensor has no mechanical wear, fast response speed, high reliability, good stability, and long service life.

2. O2 gas concentration detection adopts imported zirconia oxygen sensor, which has the characteristics of good linearity, accurate detection, and long service life.

3. Temperature inspection using PT100 temperature sensor, stable performance, good linearity. With independent door temperature control.

4. When the set value of O2 gas concentration is less than 19.0%, advanced N2 gas is used. After reaching the set value of O2 concentration, CO2 gas is introduced to ensure the accuracy of CO2 gas concentration and O2 concentration.

5. When the set value of O2 gas concentration is greater than 22.3%, advanced O2 gas is used. After reaching the set value of O2 concentration, CO2 gas is introduced to ensure the accuracy of CO2 gas concentration and O2 concentration.

6. When the box door is opened, the electronic valve automatically closes to stop the entry of any gas; the breeze circulation automatically stops, reducing the loss of N2 gas, saving gas sources, and reducing pollution caused by outside air entering the box.

7. The air circulation mode is formed from bottom to top by the fan cover and the baffles on both sides in the working room, and the breeze circulation mode is adopted to make the air circulation close to the natural air convection, shorten the recovery time of temperature, humidity and CO2 concentration, and ensure the balance of temperature, humidity, CO2 concentration and O2 concentration.

8. A separate door temperature control system allows the constant temperature control system in the box to be minimally affected by changes in ambient temperature.

9. The monitor adopts a 4.3-inch touch screen, which can display the temperature, humidity, carbon dioxide concentration and oxygen concentration in the box in real time. The control part adopts PLC control mode.

10. With temperature, carbon dioxide concentration, oxygen concentration alarm function, when the temperature exceeds the alarm value can automatically cut off all heating power.

11. With electronic valve protection function. If the CO2 intake cannot be extinguished within 30 minutes of the timer, enter the electronic valve to protect and alarm, and cut off the power supply of the CO2 intake electronic valve at the same time to protect the electronic valve from damage due to long-term power supply when no gas enters.

12. The water tray is naturally evaporated and humidified, and the humidity reaches 95% to 98%.

 

Technical parameters:


Technical indicators

Air sleeve three gas incubator

 

DM-E50SQ

DM-E80SQ

DM-E160SQ

DM-E180SQ

DM-E250SQ

 

Temperature control range

Room temperature+5℃~60℃

Constant temperature control error

±0.2℃

Temperature display error

±0.2℃

temperature fluctuation

± 0.2 ° C (measured up to ± 0.1 ° C)

temperature uniformity

± 0.5 ° C (measured up to ± 0.1 ° C)

CO2 concentration control range

0~20.0%

CO2 concentration control error

±0.2℃

CO2 concentration fluctuation

±0.2℃

O2 concentration control range

Hypoxia control range1.0~ 19.0, high oxygen control range 23.0~ 50.0 (up to 90.0%)

O2 concentration control error

± 0.2 ° C (± 0.3% outside the control range)

O2 concentration fluctuation

± 0.2 ° C (± 0.3% outside the control range)

When CO2 is set at 5.0%, the time required to rise from 0 to the set value is less than 10 minutes, the O2 concentration is set at 1.0%, and the time required to fall to the set value is less than 30 minutes

display screen

4.3-Inch touch screen (optional 7-inch)

control module

PLC

Sterilization method

UV sterilization (optional:90 degree humid heat sterilization, 180 degree dry heat sterilization)

Oxygen probe

electrochemistry

Carbon dioxide probe

Infrared probe

filter

The air inlet is equipped with a gas filter

Temperature probe

PT100

Studio sizemm

340*340*450

400*400*500

500*500*650

500*500*750

600*560*750

Air jacket volume

50L

80L

160L

180L

252L

power(±10%)

250VA

350VA

450VA

600VA

900VA

Standard configuration

Shelf2 pieces

Shelf2 pieces

Shelf3 pieces

Customer service